Precision Calibration
Calibration Management: Precision Calibration Workflow
- Precision calibration is applicable to manual calibration only.
- Precision calibration is not available for discrete and functional test calibrations.
In the following workflow diagram, the blue text in a shape indicates that the corresponding description has been provided in the sections that follow the diagram.

Administration
Persona: Analyst
Select the strategy for the template type to specify the strategy that will be used in a Calibration Profile.
Go to the Calibration Administration workflow.
Calibration Planning
Persona: Technician
Go to the Calibration Planning workflow.
Create Precision Calibration Event
Persona: Technician
If the precision calibration is enabled in the Calibration Admin Preferences page, create a Precision Calibration Event to store the results of the calibration.
Perform As Found Calibration
Persona: Technician
Enter the readings for each test point. Repeat the steps for all the measurement sets.
AF Calibration Passed?
Persona: Technician
In the Calibration record, determine if the As Found calibration has passed. This is determined based on both trueness and precision, whose results appear in the Summary and Precision Results sections respectively.
.
Recalibrate
Persona: Technician
If the calibration fails, you must recalibrate the asset.
Perform As Left Calibration
Persona: Technician
Enter the readings for each test point again after recalibrating the equipment. Repeat the steps for all the measurement sets.
AL Calibration Passed?
Persona: Technician
In the Calibration record, determine if the As Left calibration has passed. This is determined based on both trueness and precision, whose results appear in the Summary and Precision Results sections respectively.
Replace or Repair Equipment
Persona: Technician
If the As Left calibration fails, then you must replace or repair the asset before recalibrating it again.
Create Recommendation
Persona: Technician
You can create a recommendation to suggest ways to improve the calibration results or to prevent overt failures such as broken electrical conduit, failure of a protective element, rusted wires, or the presence of excess moisture.
Manage Performance Recommendation
Persona: Technician, Analyst
Manage performance recommendation is a subprocess managed outside of calibration management module.
Go to the Action Management: Manage Recommended Actions Workflow.
Approve and Close Calibration Events
Persona: Analyst
Verify the results of the calibration, and then close the calibration. After you close the calibration, all the other fields in the Calibration are disabled, and the associated Calibration Task is cycled.